Total hydrocabon analyzer for N2, Ar, He, and Air General characteristics
- Continuous measurement system based on a flame ionisation detector.
- System controlled by microprocessor.
- Easy to use .
- Easy maintenance.
- Weak sample flow.
- Less power consumption.
- CE Analyzer. Alphanumeric screen 8 x 40 characters, 64 x 240 pixels
- Multifunctional and real-time software.
Dimensions
- Height .....4U
- Width.....19'
- Depth.... 450 mm
Technical characteristics
- 0-1 ppm...display resolution 0.01 ppm
- 0-10 ppm ... display resolution 0.1 ppm
- 0-100 ppm ... display resolution 0.1 ppm
Accuracy
- 0.5% of the reading scale
- Detection limit < 0.05 ppm on the range 0-1 ppm
- Drift:... 1 % during 24 hours
- (0.5 % over a longer period)
Gas connections
- Sample.... 1 /8" Swagelok SS
- Air ... 1/8" Swagelok SS
- H 2 ... 1/8" Swagelok SS
- Sample flow ... ± 3 l/h
- Max Input Pressure... working pressure from 0.3 to 15 Bar
Synthetic air
- Flow... 20 l/h
- Pressure... 2 Bar
Hydrogen
- Flow...2 l/h
- Pressure..... 1 Bar
- Quality...6.0
- Working Temperature ... ambient and stable temperature (± 20°C)
- Power supply...220 Vac, 50-60 Hz
- Consumption...500 VA
- Output 4-20 mA... 1 output for THC impurity
- Connector RJ-45 ... computing maintenance of the system
- Output relays ... (dry contacts, SPST 2A / 250 Vac)
- 1 STATUS contact relay (security alarm)
- 3 RANGE contact relays (active range)
- 1 process alarm contact relay (level 1)
- 1 process alarm contact relay (level 2)
- 2 contacts for automatic calibration